Radiation dose is reduced with a single-pass whole-body multi-detector row CT trauma protocol compared with a conventional segmented method: initial experience.
Publication/Presentation Date
12-1-2003
Abstract
Radiation dose data were collected from a calibrated multi-detector row computed tomographic (CT) scanner during trauma CT. One protocol (used with 10 case subjects) involved a single-pass continuous whole-body acquisition from cranial vertex to symphysis pubis, while the other, conventional protocol (used with 10 control subjects) involved scouting and scanning body segments (head, cervical spine, chest, abdomen, and pelvis) individually. Technical factors were kept constant within each body segment for the single-pass and the segmented protocols. Statistics included univariate analysis, two-tailed t testing to evaluate statistical significance of the summary statistic, and power and subject population contingency tables. The mean dose length product (DLP) with the single-pass protocol was 17% lower than the sum of the DLPs of each of the individual body segment scans (P
Volume
229
Issue
3
First Page
902
Last Page
905
ISSN
0033-8419
Published In/Presented At
Ptak, T., Rhea, J. T., & Novelline, R. A. (2003). Radiation dose is reduced with a single-pass whole-body multi-detector row CT trauma protocol compared with a conventional segmented method: initial experience. Radiology, 229(3), 902–905. https://doi.org/10.1148/radiol.2293021651
Disciplines
Diagnosis | Medicine and Health Sciences | Other Analytical, Diagnostic and Therapeutic Techniques and Equipment | Radiology
PubMedID
14657320
Department(s)
Department of Radiology and Diagnostic Medical Imaging
Document Type
Article