STUDY DESIGN: Retrospective cohort study.
OBJECTIVES: To evaluate perioperative outcomes, accuracy of cage placement and radiation exposure in lateral lumbar interbody fusion (LLIF) using 3D intraoperative navigation (ION), compared to conventional 2D fluoroscopy only.
METHODS: The perioperative outcomes and accuracy of cage placement were examined in all patients who underwent LLIF using ION (ION group) or fluoroscopy only (non-ION group) by a single surgeon. The radiation exposure was examined in patients who underwent stand-alone LLIF.
RESULTS: A total of 87 patients with 154 levels (ION 49 patients with 79 levels/ non-ION 38 patients with 75 levels) were included. There were no significant differences in operative time (ION 143.5 min vs. non-ION 126.0 min,
CONCLUSIONS: The perioperative outcomes, accuracy of cage placement and radiation exposure in LLIF using ION were comparable to those using fluoroscopy only. The use of ION in LLIF was feasible, safe and accurate and may reduce radiation dose to the surgeon and surgical team.
Volume
13
Issue
3
First Page
737
Last Page
744
ISSN
2192-5682
Published In/Presented At
Urakawa, H., Sivaganesan, A., Vaishnav, A. S., Sheha, E., & Qureshi, S. A. (2023). The Feasibility of 3D Intraoperative Navigation in Lateral Lumbar Interbody Fusion: Perioperative Outcomes, Accuracy of Cage Placement and Radiation Exposure. Global spine journal, 13(3), 737–744. https://doi.org/10.1177/21925682211006700
Disciplines
Business Administration, Management, and Operations | Health and Medical Administration | Management Sciences and Quantitative Methods
PubMedID
33906453
Department(s)
Administration and Leadership
Document Type
Article