The Feasibility of 3D Intraoperative Navigation in Lateral Lumbar Interbody Fusion: Perioperative Outcomes, Accuracy of Cage Placement and Radiation Exposure.

Publication/Presentation Date

4-1-2023

Abstract

STUDY DESIGN: Retrospective cohort study.

OBJECTIVES: To evaluate perioperative outcomes, accuracy of cage placement and radiation exposure in lateral lumbar interbody fusion (LLIF) using 3D intraoperative navigation (ION), compared to conventional 2D fluoroscopy only.

METHODS: The perioperative outcomes and accuracy of cage placement were examined in all patients who underwent LLIF using ION (ION group) or fluoroscopy only (non-ION group) by a single surgeon. The radiation exposure was examined in patients who underwent stand-alone LLIF.

RESULTS: A total of 87 patients with 154 levels (ION 49 patients with 79 levels/ non-ION 38 patients with 75 levels) were included. There were no significant differences in operative time (ION 143.5 min vs. non-ION 126.0 min,

CONCLUSIONS: The perioperative outcomes, accuracy of cage placement and radiation exposure in LLIF using ION were comparable to those using fluoroscopy only. The use of ION in LLIF was feasible, safe and accurate and may reduce radiation dose to the surgeon and surgical team.

Volume

13

Issue

3

First Page

737

Last Page

744

ISSN

2192-5682

Disciplines

Business Administration, Management, and Operations | Health and Medical Administration | Management Sciences and Quantitative Methods