Failure of the Pipeline Embolization Device in Posterior Communicating Artery Aneurysms Associated with a Fetal Posterior Cerebral Artery.
Publication/Presentation Date
1-1-2016
Abstract
The pipeline embolization device has emerged as an important endovascular option. This is in part due to safety, efficacy, and possibly the ability to shorten the operative time considerably. With this new technology, some limitations are emerging as experience accumulates. We report three cases of posterior communicating (PCOM) artery aneurysms associated with fetal posterior cerebral artery where pipeline embolization was unsuccessful in obliterating the aneurysms. PCOM artery aneurysms associated with a fetal PCA should be managed either by microsurgical clipping or coiling when feasible.
Volume
2016
First Page
4691275
Last Page
4691275
ISSN
2090-6986
Published In/Presented At
Zanaty, M., Chalouhi, N., Starke, R. M., Jabbour, P., Ryken, K. O., Bulsara, K. R., & Hasan, D. (2016). Failure of the Pipeline Embolization Device in Posterior Communicating Artery Aneurysms Associated with a Fetal Posterior Cerebral Artery. Case reports in vascular medicine, 2016, 4691275. https://doi.org/10.1155/2016/4691275
Disciplines
Business Administration, Management, and Operations | Health and Medical Administration | Management Sciences and Quantitative Methods