Ventricular Perforation During Open Surgical Balloon Expandable Mitral Valve Replacement: A Case Report.

Publication/Presentation Date

11-15-2017

Abstract

A 79-year-old woman with severe mitral annular calcification was scheduled for mitral valve replacement. A SAPIEN 3 valve was implanted in mitral position using an open surgical approach. Immediately after cardiopulmonary bypass, bleeding from an unidentified source was encountered. Cardiopulmonary bypass was emergently resumed and a laceration of the left ventricular apex due to the valve delivery system was detected. Risk factors specific to the open surgical approach include a decompressed ventricle, decreased annulus to apical distance, and the absence of continuous fluoroscopic and echocardiographic imaging. These create a clinical scenario where risk of ventricular perforation is increased compared with traditional intravascular transcatheter valve delivery.

Volume

9

Issue

10

First Page

300

Last Page

304

ISSN

2325-7237

Disciplines

Business Administration, Management, and Operations | Health and Medical Administration | Management Sciences and Quantitative Methods

PubMedID

28697033

Department(s)

Administration and Leadership

Document Type

Article

Share

COinS