Orifice Areas of Balloon-Expandable Transcatheter Heart Valves: A Three-Dimensional Transesophageal Echocardiographic Study.

Publication/Presentation Date

5-1-2022

Abstract

BACKGROUND: Accurate expected effective orifice area (EOA) values for balloon-expandable (BE) transcatheter heart valves (THV) are crucial for preventing patient-prosthesis mismatch (PPM) and assessment of THV function. Currently published reference EOAs, however, are based on transthoracic echocardiography (TTE), which may be subject to left ventricular outflow tract diameter underestimation and/or suboptimal THV Doppler interrogation. The objective of this study was to establish reference EOA values for BE THVs on the basis of Doppler and three-dimensional (3D) transesophageal echocardiography (TEE).

METHODS: Two hundred twelve intraprocedural transesophageal echocardiographic examinations performed during BE THV implantation with optimal postimplantation Doppler and 3D imaging were retrospectively reviewed. Continuity equation-derived EOAs were compared with geometric orifice areas by 3D planimetry (GOA

RESULTS: The average EOA for all BE THV valves (77% SAPIEN 3) was 2.3 ± 0.5 cm

CONCLUSIONS: EOAs for BE THVs based on intraprocedural Doppler and 3D TEE suggest that previously published TTE-based reference values for EOA are underestimated, while PPM rates may be overestimated. Our findings have important clinical implications for preimplantation decision-making and for the evaluation of THV hemodynamics and function during follow-up.

Volume

35

Issue

5

First Page

460

Last Page

468

ISSN

1097-6795

Disciplines

Business Administration, Management, and Operations | Health and Medical Administration | Management Sciences and Quantitative Methods

PubMedID

34954049

Department(s)

Administration and Leadership

Document Type

Article

Share

COinS