CT based 3D printing is superior to transesophageal echocardiography for pre-procedure planning in left atrial appendage device closure.
Publication/Presentation Date
5-1-2018
Abstract
Accurate assessment of the left atrial appendage (LAA) is important for pre-procedure planning when utilizing device closure for stroke reduction. Sizing is traditionally done with transesophageal echocardiography (TEE) but this is not always precise. Three-dimensional (3D) printing of the LAA may be more accurate. 24 patients underwent Watchman device (WD) implantation (71 ± 11 years, 42% female). All had complete 2-dimensional TEE. Fourteen also had cardiac computed tomography (CCT) with 3D printing to produce a latex model of the LAA for pre-procedure planning. Device implantation was unsuccessful in 2 cases (one with and one without a 3D model). The model correlated perfectly with implanted device size (R
Volume
34
Issue
5
First Page
821
Last Page
831
ISSN
1875-8312
Published In/Presented At
Obasare, E., Mainigi, S. K., Morris, D. L., Slipczuk, L., Goykhman, I., Friend, E., Ziccardi, M. R., & Pressman, G. S. (2018). CT based 3D printing is superior to transesophageal echocardiography for pre-procedure planning in left atrial appendage device closure. The international journal of cardiovascular imaging, 34(5), 821–831. https://doi.org/10.1007/s10554-017-1289-6