Recurrent orthostatic syncope due to left atrial and left ventricular collapse after a continuous-flow left ventricular assist device implantation.
Publication/Presentation Date
1-1-2013
Abstract
Left ventricular assist devices (LVADs) have become an established treatment for patients with advanced heart failure as a bridge to transplantation or for permanent support as an alternative to heart transplantation. Continuous-flow LVADs have been shown to improve outcomes, including survival, and reduce device failure compared with pulsatile devices. Although LVADs have been shown to be a good option for patients with end-stage heart failure, unanticipated complications may occur. We describe dynamic left atrial and left ventricular chamber collapse related to postural changes in a patient with a recent continuous-flow LVAD implantation.
Volume
32
Issue
1
First Page
129
Last Page
133
ISSN
1557-3117
Published In/Presented At
Chandra, A., Pradhan, R., Kim, F. Y., Frisch, D. R., Bogar, L. J., Bonita, R., Cavarocchi, N. C., Greenspon, A. J., Hirose, H., Pitcher, H. T., Rubin, S., & Mather, P. J. (2013). Recurrent orthostatic syncope due to left atrial and left ventricular collapse after a continuous-flow left ventricular assist device implantation. The Journal of heart and lung transplantation : the official publication of the International Society for Heart Transplantation, 32(1), 129–133. https://doi.org/10.1016/j.healun.2012.10.002
Disciplines
Business Administration, Management, and Operations | Health and Medical Administration | Management Sciences and Quantitative Methods