Development of a contrast-enhanced ultrasound-guided high-intensity focused ultrasound system for coagulation of liver parenchyma.
Publication/Presentation Date
4-1-2025
Abstract
BACKGROUND: The liver is the most common organ injured in blunt abdominal trauma and makes up roughly 5% of all trauma admissions. Current treatments are invasive and resource intensive, which may delay care. We aim to develop and validate a contrast-enhanced ultrasound (CEUS)-guided noninvasive tool to treat liver lacerations at the bedside.
METHODS: Two 1.8-MHz high-intensity focused ultrasound (HIFU) elements were coupled to a C1-6 diagnostic ultrasound probe and a Logiq E10 scanner (GE HealthCare, Waukesha, WI) using a custom enclosure for coregistered imaging and ablation. A phantom was created from polyacrylamide gel combined with thermochromic ink whose color changes above biological ablative temperatures (60°C). The HIFU wave was focused approximately 0.5 cm below the surface using a 50% duty cycle generating 11.9 MPa for 20, 30, 40, 50, and 60 seconds. Experiments were repeated on ex vivo chicken livers in a water bath. Finally, the livers of four live swine underwent up to six CEUS-guided treatments using parameters optimized from in vitro work.
RESULTS: Treatment of the phantom between 20 and 60 seconds produced ablation sizes from 0.016 to 0.4 cm 3 . The relationship between time and size was exponential ( R2 = 0.992). Ablation areas were also well visualized on with ultrasound imaging. The ex vivo liver ablation size at 20 seconds was 0.37 cm 3 , at 30 seconds was 0.66 cm 3 , and at 100 seconds was 5.0 cm 3 . For the in vivo swine experiments, the average ablation area measured 2.0 × 0.75 cm with a maximum of 3.5 × 1.5 cm. Contrast-enhanced ultrasound was used with the contrast agent Definity (Lantheus Medical Imaging, North Billerica, MA) for identification of lacerations and immediate postoperative evaluation of therapy.
CONCLUSION: These experiments demonstrate the feasibility of CEUS-guided transdermal HIFU ablation and the time-dependent size of ablation. This work warrants future investigations into using ultrasound to detect active bleeding and HIFU to coagulate grades III and IV liver laceration.
Volume
98
Issue
4
First Page
662
Last Page
666
ISSN
2163-0763
Published In/Presented At
Tam, A., Contreras, K., Fall, F., Maxwell, A., Liu, J. B., Forsberg, F., Vlaisavljevich, E., Goldberg, A., Xiao, T. S., Kuon Yeng Escalante, C., Eisenbrey, J. R., & Koenig, G. (2025). Development of a contrast-enhanced ultrasound-guided high-intensity focused ultrasound system for coagulation of liver parenchyma. The journal of trauma and acute care surgery, 98(4), 662–666. https://doi.org/10.1097/TA.0000000000004397
Disciplines
Business Administration, Management, and Operations | Health and Medical Administration | Management Sciences and Quantitative Methods
PubMedID
38750641
Department(s)
Administration and Leadership
Document Type
Article