Ventricular tachycardia induced by a backup pacing stimulus in a patient with a dual chamber pacemaker with AutoCapture.
Publication/Presentation Date
6-1-2021
Abstract
Manufacturers of cardiac implantable electronic devices have incorporated automatic features to allow for remote monitoring, improve device longevity, and additional safety. Algorithms to automatically measure capture threshold and adjust output to preserve battery life are one such feature. Automatic features may occasionally result in unexpected or undesirable clinical outcomes. We report on a patient who developed ventricular tachycardia inadvertently induced by the AutoCapture. feature of an Abbott/St. Jude Medical (SJM) pacemaker.
Volume
44
Issue
6
First Page
1130
Last Page
1132
ISSN
1540-8159
Published In/Presented At
Greenspon, A. J., Frisch, D. R., Ho, R. T., & Pavri, B. B. (2021). Ventricular tachycardia induced by a backup pacing stimulus in a patient with a dual chamber pacemaker with AutoCapture. Pacing and clinical electrophysiology : PACE, 44(6), 1130–1132. https://doi.org/10.1111/pace.14199