Dual atrioventricular nodal pathways physiology and high-grade heart block.

Publication/Presentation Date

1-1-2020

Abstract

We report of case of an 87 year old lady with preexisting RBBB who developed LBBB after transcatheter aortic valve replacement (TAVR) for the treatment of severe aortic stenosis. She underwent pacemaker implantation, and subsequently developed high-grade atrioventricular (AV) block. Dual chamber pacing in the setting of complete heart block with a long programmed AV delay showed retrograde P waves. Ventricular pacing showed intact retrograde conduction. Shortening the programmed AV delay resulted in loss of retrograde P waves during dual chamber pacing. These findings are discussed.

Volume

62

First Page

138

Last Page

141

ISSN

1532-8430

Disciplines

Business Administration, Management, and Operations | Health and Medical Administration | Management Sciences and Quantitative Methods

PubMedID

32866914

Department(s)

Administration and Leadership

Document Type

Article