Contained rupture of a pseudoaneurysm of the descending thoracic aorta related to remnant outflow graft of left ventricular assist device after heart transplantation.
Publication/Presentation Date
10-1-2012
Abstract
The Jarvik 2000 left ventricular assist device is inserted via a left thoracotomy with the outflow graft anastomosed to the descending thoracic aorta. Removal of the device during heart transplantation involves division of the outflow graft, resulting in a retained remnant. We describe the first reported case of a mycotic pseudoaneurysm of the descending thoracic aorta related to the remnant of a left ventricular assist device outflow graft in an immunosuppressed heart recipient complicated with systemic Pseudomonas infection. The pseudoaneurysm was temporarily treated with endovascular stent grafting followed by delayed thoracotomy, pseudoaneurysm excision, and placement of an aortic interposition graft using an aortic allograft.
Volume
94
Issue
4
First Page
1345
Last Page
1348
ISSN
1552-6259
Published In/Presented At
Yamane, K., Bogar, L. J., DiMuzio, P. J., Cowan, S. W., Hirose, H., Evans, N. R., 3rd, Rao, A. S., Eisenberg, J. A., & Cavarocchi, N. C. (2012). Contained rupture of a pseudoaneurysm of the descending thoracic aorta related to remnant outflow graft of left ventricular assist device after heart transplantation. The Annals of thoracic surgery, 94(4), 1345–1348. https://doi.org/10.1016/j.athoracsur.2012.02.023
Disciplines
Business Administration, Management, and Operations | Health and Medical Administration | Management Sciences and Quantitative Methods
PubMedID
23006696
Department(s)
Administration and Leadership
Document Type
Article