Prospective, multicenter, international phase 2 trial evaluating ultrasonic energy for pulmonary artery branch sealing in video-assisted thoracoscopic surgery lobectomy.

Publication/Presentation Date

1-1-2020

Abstract

OBJECTIVES: The study objectives were to evaluate the immediate, short-, and medium-term efficacy and safety of pulmonary artery branch sealing using an ultrasonic vessel-sealing device in minimally invasive anatomic lung resection.

METHODS: This study consists of a prospective, phase 2, multicenter, international clinical trial (clinicaltrials.gov: NCT02719717) that enrolled patients planned for video-assisted thoracoscopic surgery/robotic anatomic lung resection in 7 centers (United States, Canada, United Kingdom). Pulmonary artery branches of 7 mm or less were sealed and divided with an ultrasonic energy vessel-sealing device. The remainder of the lobectomy was performed according to surgeon preference. Intraoperative, in-hospital, and 30-day postoperative bleeding and complications were prospectively recorded.

RESULTS: A total of 150 patients with a minimum of 1 pulmonary artery branch sealed with an ultrasonic vessel-sealing device were prospectively enrolled in the trial. Resections included 139 lobectomies and 11 segmentectomies. A total of 424 pulmonary artery branches were divided: 239 with the ultrasonic vessel-sealing device, 181 with endostaplers, and 4 with endoscopic clips. The mean and median pulmonary artery diameters were 4.7 mm/5.0 mm, 10.3 mm/10.0 mm, and 6.5 mm/6.5 mm for each method, respectively. Three of the pulmonary artery branches divided with the ultrasonic vessel-sealing device (1.3%) and 4 pulmonary artery branches divided with endostaplers (2.2%) bled intraoperatively. Among the patients with seal failures, 1 patient required conversion to thoracotomy. There was no postoperative bleeding from divided pulmonary artery branches with either sealing method. There was no mortality at 30 days.

CONCLUSIONS: Pulmonary artery branch sealing with ultrasonic energy during video-assisted thoracoscopic surgery lobectomy is safe for vessels 7 mm or less. The use of an ultrasonic device is a reasonable sealing method for pulmonary artery branches 7 mm or less.