Clinical test cases for commissioning, QA, and benchmarking of model-based dose calculation algorithms in

Publication/Presentation Date

1-1-2026

Abstract

PURPOSE: To develop clinically relevant test cases for commissioning Model-Based Dose Calculation Algorithms (MBDCAs) for

ACQUISITION AND VALIDATION METHODS: Two cervical cancer intracavitary HDR brachytherapy models were developed based on a real patient, using either uniformly structured regions or realistic segmentation. The patient's computed tomography (CT) images were processed, converted to a series of digital imaging and communications in medicine (DICOM) CT images, and imported into two treatment planning systems (TPSs), the Oncentra Brachy and BrachyVision. The original segmentation of the clinical case was augmented to enable a thorough dosimetric analysis. The actual clinical treatment plan was generally maintained, with the source replaced by a generic

DATA FORMAT AND USAGE NOTES: The datasets for the developed cases are available online at https://doi.org/10.5281/zenodo.15720996. The DICOM files include the treatment plan for each case, TPS, and the corresponding reference MC dose data. The package also contains a TPS- and case-specific user guide for commissioning the MBDCAs, as well as files necessary to replicate the MC simulations.

POTENTIAL APPLICATIONS: The provided datasets and proposed methodology can serve as a commissioning framework for TPSs that employ MBDCAs, as well as a benchmark for brachytherapy researchers using MC methods and MBDCA developers. They also facilitate intercomparisons of MBDCA performance and provide a quality assurance resource for evaluating future TPS software updates.

Volume

53

Issue

1

First Page

70260

Last Page

70260

ISSN

2473-4209

Disciplines

Business Administration, Management, and Operations | Health and Medical Administration | Management Sciences and Quantitative Methods