Clinical transition to model-based dose calculation algorithm: A retrospective analysis of high-dose-rate tandem and ring brachytherapy of the cervix.

Publication/Presentation Date

1-1-2017

Abstract

PURPOSE: To retrospectively compare clinical dosimetry of CT-based tandem-ring treatment plans using a model-based dose calculation algorithm (MBDCA) with the standard TG-43-based dose formalism.

METHODS AND MATERIALS: A cohort of 10 cervical cancer cohorts treated using the tandem and ring high-dose-rate applicators were evaluated. The original treatment plans were created using the department CT-based volume optimization clinical standards. All plans originally calculated with TG-43 dose calculation formalism were recalculated using the MBDCA algorithm. The gross target volume and organs at risk (OARs) were contoured on each data set along with significant heterogeneities like air in cavity and high-density plastic tandem and ring components. The patient tissue was modeled as homogenous liquid water. D

RESULTS: Using the manual applicator contouring method, MBDCA plans (n = 10) showed 2.1 ± 1.1% reduction in dose to Point A average, 2.6 ± 0.9% reduction in Target D

CONCLUSION: For unshielded plastic gynecologic applicators, minimal dosimetric changes (< 5%) were found using MBDCA relative to standard TG-43. Use of solid applicator model is more efficient than manual applicator contouring and also yielded similar MBDCA dosimetric results. Currently, TG-186 dose calculations should be reported along TG-43 until we obtain studies with larger cohorts to fully realize the potential of MBDCA dosimetry.

Volume

16

Issue

3

First Page

624

Last Page

629

ISSN

1873-1449

Disciplines

Business Administration, Management, and Operations | Health and Medical Administration | Management Sciences and Quantitative Methods

PubMedID

28365083

Department(s)

Administration and Leadership

Document Type

Article

Share

COinS