A generic TG-186 shielded applicator for commissioning model-based dose calculation algorithms for high-dose-rate
Publication/Presentation Date
11-1-2017
Abstract
PURPOSE: A joint working group was created by the American Association of Physicists in Medicine (AAPM), the European Society for Radiotherapy and Oncology (ESTRO), and the Australasian Brachytherapy Group (ABG) with the charge, among others, to develop a set of well-defined test case plans and perform calculations and comparisons with model-based dose calculation algorithms (MBDCAs). Its main goal is to facilitate a smooth transition from the AAPM Task Group No. 43 (TG-43) dose calculation formalism, widely being used in clinical practice for brachytherapy, to the one proposed by Task Group No. 186 (TG-186) for MBDCAs. To do so, in this work a hypothetical, generic high-dose rate (HDR)
METHODS: A generic HDR
RESULTS: The local dose differences among MC codes are comparable to the statistical uncertainties of the reference datasets for the "source centered in water" and "source displaced" test cases and for the clinically relevant part of the unshielded volume in the "source centered in applicator" case. Larger local differences appear in the shielded volume or at large distances. Considering clinically relevant regions, global dose differences are smaller than the local ones. The most disadvantageous case for the MBDCAs is the one including the shielded applicator. In this case, ACUROS agrees with MC within [-4.2%, +4.2%] for the majority of voxels (95%) while presenting dose differences within [-0.12%, +0.12%] of the dose at a clinically relevant reference point. For ACE, 95% of the total volume presents differences with respect to MC in the range [-1.7%, +0.4%] of the dose at the reference point.
CONCLUSIONS: The combination of the generic source and generic shielded applicator, together with the previously developed test cases and reference datasets (available in the Brachytherapy Source Registry), lay a solid foundation in supporting uniform commissioning procedures and direct comparisons among treatment planning systems for HDR
Volume
44
Issue
11
First Page
5961
Last Page
5976
ISSN
2473-4209
Published In/Presented At
Ma, Y., Vijande, J., Ballester, F., Tedgren, Å. C., Granero, D., Haworth, A., Mourtada, F., Fonseca, G. P., Zourari, K., Papagiannis, P., Rivard, M. J., Siebert, F. A., Sloboda, R. S., Smith, R., Chamberland, M. J. P., Thomson, R. M., Verhaegen, F., & Beaulieu, L. (2017). A generic TG-186 shielded applicator for commissioning model-based dose calculation algorithms for high-dose-rate 192 Ir brachytherapy. Medical physics, 44(11), 5961–5976. https://doi.org/10.1002/mp.12459
Disciplines
Business Administration, Management, and Operations | Health and Medical Administration | Management Sciences and Quantitative Methods
PubMedID
28722180
Department(s)
Administration and Leadership
Document Type
Article