A MC-based anthropomorphic test case for commissioning model-based dose calculation in interstitial breast 192-Ir HDR brachytherapy.

Publication/Presentation Date

7-1-2023

Abstract

PURPOSE: To provide the first clinical test case for commissioning of

ACQUISITION AND VALIDATION METHODS: A computational patient phantom model was generated from a clinical multi-catheter

DATA FORMAT AND USAGE NOTES: The dataset is available online at http://irochouston.mdanderson.org/rpc/BrachySeeds/BrachySeeds/index.html,https://doi.org/10.52519/00005. Files include the treatment plan for each TPS in DICOM RT format, reference MC dose data in RT Dose format, as well as a guide for database users and all files necessary to repeat the MC simulations.

POTENTIAL APPLICATIONS: The dataset facilitates the commissioning of brachytherapy MBDCAs using TPS embedded tools and establishes a methodology for the development of future clinical test cases. It is also useful to non-MBDCA adopters for intercomparing MBDCAs and exploring their benefits and limitations, as well as to brachytherapy researchers in need of a dosimetric and/or a DICOM RT information parsing benchmark. Limitations include specificity in terms of radionuclide, source model, clinical scenario, and MBDCA version used for its preparation.

Volume

50

Issue

7

First Page

4675

Last Page

4687

ISSN

2473-4209

Disciplines

Business Administration, Management, and Operations | Health and Medical Administration | Management Sciences and Quantitative Methods

PubMedID

37194638

Department(s)

Administration and Leadership

Document Type

Article

Share

COinS