Reconstruction of volume averaging effect-free continuous photon beam profiles from discrete ionization chamber array measurements using a machine learning technique.

Publication/Presentation Date

10-1-2021

Abstract

PURPOSE: The use of the ionization chamber array ICProfiler (ICP) is limited by its relatively poor detector spatial resolution and the inherent volume averaging effect (VAE). The purpose of this work is to study the feasibility of reconstructing VAE-free continuous photon beam profiles from ICP measurements with a machine learning technique.

METHODS: In- and cross-plane photon beam profiles of a 6 MV beam from an Elekta linear accelerator, ranging from 2 × 2 to 10 × 10 cm

RESULTS: The profiles reconstructed with all the ANNs had excellent agreement with the reference. For in-plane, the ANNs reduced the PWD from 1.6 ± 0.7 mm at 1.5 cm depth to 0.1 ± 0.1 mm, from 1.8 ± 0.6 mm at 5.0 cm depth to 0.1 ± 0.1 mm, and from 2.4 ± 0.1 mm at 10.0 cm depth to 0.0 ± 0.0 mm; for cross-plane, the ANNs reduced the PWD from 1.2 ± 0.4 mm at 1.5 cm depth, 1.2 ± 0.3 mm at 5.0 cm depth, and 1.6 ± 0.1 mm at 10.0 cm depth, to 0.1 ± 0.1 mm.

CONCLUSIONS: This study demonstrated the feasibility of using simple ANNs to reconstruct VAE-free continuous photon beam profiles from discrete ICP measurements. A combined ANN can restore the penumbra of in- and cross-plane beam profiles of various fields at different depths.

Volume

22

Issue

10

First Page

161

Last Page

168

ISSN

1526-9914

Disciplines

Business Administration, Management, and Operations | Health and Medical Administration | Management Sciences and Quantitative Methods

PubMedID

34486800

Department(s)

Administration and Leadership