Evaluation of a neural network-based photon beam profile deconvolution method.

Publication/Presentation Date

6-1-2020

Abstract

PURPOSE: The authors have previously shown the feasibility of using an artificial neural network (ANN) to eliminate the volume average effect (VAE) of scanning ionization chambers (ICs). The purpose of this work was to evaluate the method when applied to beams of different energies (6 and 10 MV) and modalities [flattened (FF) vs unflattened (FFF)], measured with ICs of various sizes.

METHODS: The three-layer ANN extracted data from transverse photon beam profiles using a sliding window, and output deconvolved value corresponding to the location at the center of the window. Beam profiles of seven fields ranging from 2 × 2 to 10 × 10 cm

RESULTS: Excellent agreement between the deconvolved and reference profiles was achieved with both separate and combined ANNs for all studied ICs, beam energies, beam modalities, and geometries. After deconvolution, the average PWD decreased from 1-3 mm to under 0.15 mm with separate ANNs and to under 0.20 mm with combined ANN.

CONCLUSIONS: The ANN-based deconvolution method can be effectively applied to beams of different energies and modalities measured with ICs of various sizes. Separate ANNs yielded marginally better results than combined ANNs. An IC-specific, combined ANN can provide clinically acceptable results as long as the training data includes data of each beam energy and modality.

Volume

21

Issue

6

First Page

53

Last Page

62

ISSN

1526-9914

Disciplines

Business Administration, Management, and Operations | Health and Medical Administration | Management Sciences and Quantitative Methods

PubMedID

32227629

Department(s)

Administration and Leadership

Document Type

Article

Share

COinS