Identification of cholangiocarcinoma by using the Spyglass Spyscope system for peroral cholangioscopy and biopsy collection.

Publication/Presentation Date

5-1-2012

Abstract

BACKGROUND & AIMS: It is a challenge to collect samples from bile duct strictures to diagnose patients with cholangiocarcinoma. We investigated the utility of the Spyglass Spyscope, a single-operator endoscope that is used to perform cholangiopancreatoscopy, to identify extrahepatic cholangiocarcinoma in patients who were not diagnosed with this disorder by endoscopic retrograde cholangiopancreatography (ERCP) cytology or endoscopic ultrasound-guided fine-needle aspiration (EUS-FNA) analyses.

METHODS: We conducted a retrospective analysis of data from 30 patients (median age, 67 years; 67% male) with indeterminate extrahepatic biliary strictures who were ultimately diagnosed with cholangiocarcinoma but had inconclusive results from initial biliary ductal brush cytology and EUS-FNA analyses. Patients then underwent cholangioscopy by using the Spyglass Spyscope and intraductal biopsy analysis. None of the patients had a definitive mass in abdominal imaging or EUS analyses.

RESULTS: The biliary stricture was located in the common bile duct in 13 patients and in the common hepatic duct in 17 patients. The Spyglass Spyscope system had 77% accuracy (23 of 30) in the diagnosis of malignancies that were inconclusive on the basis of ERCP-guided brush or EUS-FNA analyses.

CONCLUSIONS: The Spyglass Spyscope for cholangioscopy and biopsy collection identified malignancies with 77% accuracy in patients with suspected cholangiocarcinoma.

Volume

10

Issue

5

First Page

466

Last Page

471

ISSN

1542-7714

Disciplines

Business Administration, Management, and Operations | Health and Medical Administration | Management Sciences and Quantitative Methods

PubMedID

22178463

Department(s)

Administration and Leadership

Document Type

Article

Share

COinS