Automated and Manual Measurements of the Aortic Annulus with ECG-Gated Cardiac CT Angiography Prior to Transcatheter Aortic Valve Replacement: Comparison with 3D-Transesophageal Echocardiography.
Publication/Presentation Date
5-1-2017
Abstract
RATIONALE AND OBJECTIVES: Multimodality evaluation of the aortic annulus is generally advocated to plan for transcatheter aortic valve replacement (TAVR). We compared aortic annular measurements by cardiac computed tomography angiography (cCTA) to three-dimensional transesophageal echocardiography (3D-TEE), and also evaluated the use of semi-automated software for cCTA annular measurements.
MATERIALS AND METHODS: A retrospective cohort of 74 patients underwent 3D-TEE and electrocardiogram-gated cCTA of the heart within 30 days for TAVR planning. 3D-TEE measurements were obtained during mid-systole; cCTA measurements were obtained during late-systole (40% of R-R interval) and mid-diastole (80% of R-R interval). Annular area was measured independently by manual planimetry and with semi-automated software.
RESULTS: cCTA measurements in systole and diastole were highly correlated for short-axis diameter (r = 0.91), long-axis diameter (r = 0.92), and annular area (r = 0.96), although systolic measurements were significantly larger (P < 0.001), most notably for the short-axis diameter. Good correlation was observed between 3D-TEE and cCTA for short-axis diameter (r = 0.84-0.90), long-axis diameter (r = 0.77-0.79), and annular area (r = 0.89-0.90). As compared to 3D-TEE, annular area is overmeasured by 28 mm
CONCLUSIONS: Although all cCTA measurements are highly correlated with measurements by 3D-TEE, diastolic phase cCTA measurements tend to be closer to standard mid-systolic 3D-TEE measurements. Semi-automated measurement of the aortic annulus with cCTA is highly correlated with manual planimetry. Nonetheless, annular contours derived by semi-automated software should be visually inspected, as the variability in area estimates for individual cases between manual and automated measurements may alter the sizing of an aortic prosthesis.
Volume
24
Issue
5
First Page
587
Last Page
593
ISSN
1878-4046
Published In/Presented At
Guez, D., Boroumand, G., Ruggiero, N. J., Mehrotra, P., & Halpern, E. J. (2017). Automated and Manual Measurements of the Aortic Annulus with ECG-Gated Cardiac CT Angiography Prior to Transcatheter Aortic Valve Replacement: Comparison with 3D-Transesophageal Echocardiography. Academic radiology, 24(5), 587–593. https://doi.org/10.1016/j.acra.2016.12.008
Disciplines
Business Administration, Management, and Operations | Health and Medical Administration | Management Sciences and Quantitative Methods
PubMedID
28130049
Department(s)
Administration and Leadership
Document Type
Article