Hydrophilic inflatable penile prosthesis surface coatings readily rebind antibiotics and maintain antimicrobial efficacy ex vivo.
Publication/Presentation Date
2-1-2026
Abstract
Hydrophilic penile prosthesis (PP) surfaces removed during revision surgery may potentially rebind antiseptics and maintain antimicrobial efficacy ex vivo. Coloplast Titan reservoirs and cylinders were retrieved during revision surgery for mechanical failure. Congo Red staining and contact angle measurements were performed to evaluate the integrity of the hydrophilic surface. Fluorescent antibiotic binding was performed by submerging discs for 3 min in either fluorescein-isothiocyanate (FITC) labeled or unlabeled vancomycin at 2 mg/mL. Fluorescence was quantified via ImageJ. 8 mm discs were submerged for 3 min in normal saline (NS), 0.05% chlorhexidine gluconate, or a 2 mg/mL vancomycin and 160 μg/mL gentamicin (VG) antibiotic, then incubated with 10
Volume
38
Issue
2
First Page
134
Last Page
140
ISSN
1476-5489
Published In/Presented At
Im, B. H., Hochberg, A. R., Giordano, A. H., Evans, R., Fertala, A., Fertala, J., Hickok, N. J., & Chung, P. H. (2026). Hydrophilic inflatable penile prosthesis surface coatings readily rebind antibiotics and maintain antimicrobial efficacy ex vivo. International journal of impotence research, 38(2), 134–140. https://doi.org/10.1038/s41443-025-01104-8
Disciplines
Business Administration, Management, and Operations | Health and Medical Administration | Management Sciences and Quantitative Methods
PubMedID
40500460
Department(s)
Administration and Leadership
Document Type