Robotic camera holder as good as expert camera holder: a randomized crossover trial.
Publication/Presentation Date
6-1-2009
Abstract
INTRODUCTION: This study aimed to compare the impact of robotic camera holder (RCH) and human camera holder (HCH) on product quality and procedure effectiveness of a simulated laparoscopic procedure.
METHODS: This was a prospective randomized crossover trial including voluntary surgical residents. Block randomization generated RCH-HCH or HCH-RCH sequence allocation. The task was suturing a duodenal perforation on foam stomach with intracorporeally knot tying in a simulator. The camera was operated by the same robot and same expert. Product quality was measured by accuracy error, tissue damage, sliding knot, and leak. Procedure effectiveness was measured by operating time, nongoal directed actions, and dangerous actions. Kendall's coefficient tau_b was used for interrater reliability between 2 blinded assessors.
RESULTS: Forty-four subjects performed their tasks as allocated. Product quality and procedure effectiveness were similar when first attempt of task was compared with the repeat task by same subject ignoring the type of camera holder. There was no evidence of significant unequal carryover effect when comparison was stratified by RCH-HCH or HCH-RCH sequences. There were no differences in product quality and procedure effectiveness when RCH was compared with HCH. Coefficient tau_b was > or = 0.80 for all but dangerous actions (0.72, P=0.08).
CONCLUSIONS: RCH and HCH had similar impact on product quality and procedure effectiveness of simulated laparoscopic procedure.
Volume
19
Issue
3
First Page
272
Last Page
275
ISSN
1534-4908
Published In/Presented At
Uchal, M., Haughn, C., Raftopoulos, Y., Hussain, F., Reed, J. F., Tjugum, J., & Bergamaschi, R. (2009). Robotic camera holder as good as expert camera holder: a randomized crossover trial. Surgical laparoscopy, endoscopy & percutaneous techniques, 19(3), 272–275. https://doi.org/10.1097/SLE.0b013e3181a6da6c
Disciplines
Anesthesiology | Medicine and Health Sciences
PubMedID
19542861
Department(s)
Department of Anesthesiology
Document Type
Article