Premarket Evaluation of IDS RapID SS/u System for Identification of Urine Isolates.
Publication/Presentation Date
1-1-1987
Abstract
A total of 170 fresh clinical urine isolates were tested with a premarket configuration of the RapID SS/u system (Innovative Diagnostic Systems, Inc., Atlanta, Ga.), a qualitative micromethod for the identification of selected organisms commonly isolated from urine specimens. Results were compared with those obtained with conventional methods of identifying gram-positive isolates and with the AutoMicrobic system (Vitek Systems, Inc., Hazelwood, Mo.), utilizing Gram-Negative Identification cards for the identification of gram-negative rods. Organisms representing 12 taxa were included in the study. Of the 170 isolates, 163 (95.9%) were correctly identified. A total of 144 strains (84.7%) were correctly identified without additional testing, whereas 19 isolates (11.2%) required further testing. Seven isolates (4.1%) were incorrectly identified. The SS/u system required minimal hands-on time inoculate and interpret reactions. Discrepancies most often occurred with regard to misinterpretation of Escherichia coli and Enterobacter sp. as Citrobacter sp. The IDS RapID SS/u system may indeed prove valuable for the rapid manual identification of urine isolates.
Volume
25
Issue
1
First Page
42
Last Page
44
ISSN
0095-1137
Published In/Presented At
Halstead, D. C., Hoffert, M. R., & Colasante, G. G. (1987). Premarket evaluation of IDS RapID SS/u system for identification of urine isolates. Journal Of Clinical Microbiology, 25(1), 42-44.
Disciplines
Medical Pathology | Pathology
PubMedID
3539997
LVHN link
http://search.ebscohost.com/login.aspx?direct=true&db=mnh&AN=3539997&site=ehost-live&scope=site
Department(s)
Department of Surgery, Department of Surgery Faculty
Document Type
Article