Assessment of the residual error in soft tissue setup in patients undergoing partial breast irradiation: results of a prospective study using cone-beam computed tomography.
Publication/Presentation Date
3-15-2008
Abstract
PURPOSE: On-board cone-beam computed tomography (CBCT) provides soft tissue information that may improve setup accuracy in patients undergoing accelerated partial breast irradiation (APBI). We used CBCT to assess the residual error in soft tissue after two-dimensional kV/MV alignment based on bony anatomy. We also assessed the dosimetric impact of this error.
METHODS AND MATERIALS: Ten patients undergoing APBI were studied as part of an institutional review board-approved prospective trial. Patients were aligned based on skin/cradle marks plus orthogonal kV/MV images registered based on bony landmarks to digitally reconstructed radiographs from the planning CT. A subsequent CBCT was registered to the planning CT using soft tissue information. This "residual error" and its dosimetric impact was measured.
RESULTS: The root-mean-square of the residual error was 3, 4, and 4 mm, in the right-left, anterior-posterior, and superior-inferior directions, respectively. The average vector sum was 6+/-2 mm. Average reductions in mean dose to the lumpectomy cavity, clinical target volume (CTV), and planning target volume were 0.1%, 0.4%, and 1%, respectively. The mean difference in the clinical target and planning target volumes that received 95% of the prescribed dose (V95) were 1% and 4%.
CONCLUSIONS: In this initial study with a modest number of patients, the residual error in soft tissue was typically <5 >mm, and with the field margins used, the resultant dosimetric consequences were modest. In patients immobilized in a customized cradle, setup using orthogonal kV images thus appears accurate and reproducible. The CBCT technique may have particular utility in patients with larger breast volumes or breast deformations. Further studies involving larger numbers of patients are needed to further assess the utility of CBCT.
Volume
70
Issue
4
First Page
1025
Last Page
1034
ISSN
0360-3016
Published In/Presented At
Fatunase, T., Wang, Z., Yoo, S., Hubbs, J. L., Prosnitz, R. G., Yin, F. F., & Marks, L. B. (2008). Assessment of the residual error in soft tissue setup in patients undergoing partial breast irradiation: results of a prospective study using cone-beam computed tomography. International journal of radiation oncology, biology, physics, 70(4), 1025–1034. https://doi.org/10.1016/j.ijrobp.2007.07.2344
Disciplines
Medicine and Health Sciences | Oncology
PubMedID
17892919
Department(s)
Department of Radiation Oncology
Document Type
Article