Early systole in the healthy kidney: variability of Doppler US waveform parameters.

Publication/Presentation Date

10-1-1997

Abstract

PURPOSE: To determine the sampling variability of intrarenal Doppler ultrasound (US) indexes of early systole in a screened population of healthy individuals.

MATERIALS AND METHODS: Doppler US measurements were obtained in the superior, middle, and inferior regions of 132 kidneys in 66 healthy, potential kidney donors. All individuals subsequently underwent angiography. Systematic measurement differences and random effects were estimated for sampling from a particular parenchymal region, from a kidney in a subject, from an individual in a population, and from a vascular territory.

RESULTS: Coefficients of variations ranged from 20% to 30%. Most (55%-66%) of the observed variation was attributable to random differences between repeated measurements in the same kidney. No systematic variations attributable to kidney region, vascular territory, right versus left kidney, or subject age were found for acceleration time, acceleration, or waveform shape. Some evidence of fixed variation between kidneys and between regions was found for peak systolic velocity, but the magnitude of this variation was small. Averages of repeated measurements may decrease the probability of exceeding the normal threshold for acceleration but not for acceleration time.

CONCLUSION: Measurements of Doppler parameters of the early systole have substantial intrinsic variability. Thus, caution is needed when interpreting small changes in these measurements within a kidney or between individuals.

Volume

205

Issue

1

First Page

109

Last Page

113

ISSN

0033-8419

Disciplines

Diagnosis | Medicine and Health Sciences | Other Analytical, Diagnostic and Therapeutic Techniques and Equipment | Radiology

PubMedID

9314971

Department(s)

Department of Radiology and Diagnostic Medical Imaging

Document Type

Article

Share

COinS