MR of cervical lymph nodes: comparison of fast spin-echo and conventional spin-echo T2W scans.

Publication/Presentation Date

10-1-1994

Abstract

PURPOSE: In order to explore the utility of fast spin-echo techniques in the neck, 50 consecutive conventional spin-echo (CSE) long TR examinations were compared with 50 consecutive fast spin-echo (FSE) long TR scans for cervical lymph nodes.

MATERIALS AND METHODS: Standard CSE examinations used parameters of 2200/80/1 (TR/TE/excitations), 256 x 128 matrix, 5 mm thick with interslice gaps between 1 and 2.5 mm. FSE studies were employed with TR of 4000/80/2, 256 x 192 matrix, fat suppression and 5 mm contiguous slices. Standard T1-weighted examinations and clinical correlation were used as proof of nodal presence.

RESULTS: Overall vascular flow artefacts, patient motion artefacts, and image quality were comparable between the two techniques. Lymph node detectability was superior with FSE scans (P < 0.05). Typical time saved was approximately 4.5 min with FSE, despite the use of larger matrices, contiguous slices and twice the excitations.

CONCLUSION: FSE is a competitive technique to CSE images in the neck, yielding similar quality images with higher resolution, reduced scan time, and improved lymph node detectability.

Volume

49

Issue

10

First Page

670

Last Page

675

ISSN

0009-9260

Disciplines

Diagnosis | Medicine and Health Sciences | Other Analytical, Diagnostic and Therapeutic Techniques and Equipment | Radiology

PubMedID

7955826