Omission of Planning CT Reduces Patient Radiation Exposure during CT-Guided Bone Marrow Biopsy and Aspiration.
Publication/Presentation Date
11-8-2021
Abstract
The purpose of this study is to evaluate the impact of eliminating a preprocedural planning computed tomography during CT-guided bone marrow biopsy on the technical aspects of the procedure, including patient dose, sample quality, procedure time, and CT fluoroscopy usage. Retrospective analysis of 109 patients between 1 June 2018 and 1 January 2021 was performed. Patients were grouped based on whether they received a planning CT scan. Relative radiation exposure was measured using dose-length product (DLP). Secondary metrics included number of CT fluoroscopic acquisitions until target localization, total number of CT fluoroscopic acquisitions, biopsy diagnostic yield, and procedure time. A total of 43 bone marrow biopsies with planning CT scans (Group 1) and 66 bone marrow biopsies without planning CT scans (Group 2) were performed. The average total DLP for Group 1 and Group 2 was 268.73 mGy*cm and 50.92 mGy*cm, respectively. The mean radiation dose reduction between the groups was 81% (
Volume
7
Issue
4
First Page
747
Last Page
751
ISSN
2379-139X
Published In/Presented At
Devita, R., Chagarlamudi, K., Durieux, J., Jordan, D., Nguyen, B., Davidson, J., Kosaraju, V., & Abboud, S. (2021). Omission of Planning CT Reduces Patient Radiation Exposure during CT-Guided Bone Marrow Biopsy and Aspiration. Tomography (Ann Arbor, Mich.), 7(4), 747–751. https://doi.org/10.3390/tomography7040062
Disciplines
Diagnosis | Medicine and Health Sciences | Other Analytical, Diagnostic and Therapeutic Techniques and Equipment | Radiology
PubMedID
34842812
Department(s)
Department of Radiology and Diagnostic Medical Imaging
Document Type