Omission of Planning CT Reduces Patient Radiation Exposure during CT-Guided Bone Marrow Biopsy and Aspiration.

Publication/Presentation Date

11-8-2021

Abstract

The purpose of this study is to evaluate the impact of eliminating a preprocedural planning computed tomography during CT-guided bone marrow biopsy on the technical aspects of the procedure, including patient dose, sample quality, procedure time, and CT fluoroscopy usage. Retrospective analysis of 109 patients between 1 June 2018 and 1 January 2021 was performed. Patients were grouped based on whether they received a planning CT scan. Relative radiation exposure was measured using dose-length product (DLP). Secondary metrics included number of CT fluoroscopic acquisitions until target localization, total number of CT fluoroscopic acquisitions, biopsy diagnostic yield, and procedure time. A total of 43 bone marrow biopsies with planning CT scans (Group 1) and 66 bone marrow biopsies without planning CT scans (Group 2) were performed. The average total DLP for Group 1 and Group 2 was 268.73 mGy*cm and 50.92 mGy*cm, respectively. The mean radiation dose reduction between the groups was 81% (

Volume

7

Issue

4

First Page

747

Last Page

751

ISSN

2379-139X

Disciplines

Diagnosis | Medicine and Health Sciences | Other Analytical, Diagnostic and Therapeutic Techniques and Equipment | Radiology

PubMedID

34842812