Occurrence and clinical implications of heterotopic ossification after cervical disc arthroplasty with the Prestige LP Cervical Disc at 2 contiguous levels.

Publication/Presentation Date

3-13-2020

Abstract

OBJECTIVE: The authors sought to assess the impact of heterotopic ossification (HO) on clinical outcomes and angular range of motion (ROM) after cervical disc arthroplasty (CDA) performed with the Prestige LP Cervical Disc (Prestige LP disc) at 2 levels.

METHODS: HO was assessed and graded from 0 to IV for increasing severity on lateral neutral radiographs at each visit in 209 patients who underwent implantation of Prestige LP discs at 2 cervical levels in a clinical trial with extended 10-year follow-up. ROM was compared by using HO grade, and clinical outcomes were compared between HO subgroups (grade 0-II vs III/IV) based on HO severity at 2 and 10 years after surgery.

RESULTS: The grade III/IV HO incidence at either or both index levels was 24.2% (48/198) at 2 years and 39.0% (57/146) at 10 years. No statistical difference was found in overall success; neurological success; or Neck Disability Index (NDI), neck pain, arm pain, or SF-36 Physical Component Summary (PCS) scores between the HO subgroups (grade 0-II vs III/IV) at either 2 or 10 years. The cumulative rate of possible implant-related adverse events (AEs) was higher in patients having grade III/IV HO at 2 years (56.3%) and 10 years (47.8%) compared with those having grade 0-II HO at 2 years (24.4%) and 10 years (17.9%), specifically in 2 subcategories: spinal events and HOs reported by the investigators. No statistical difference was found between the HO subgroups in possible implant-related serious AEs or secondary surgeries at the index or adjacent levels. The average angular ROMs at index levels were lower in subjects with higher-grade HO at 2 and 10 years. The average ROMs at the superior level were 8.8°, 6.6°, 3.2°, and 0.3°, respectively, for the HO grade 0/I, II, III, and IV groups at 10 years, and 7.9°, 6.2°, 3.7°, and 0.6°, respectively, at the inferior level.

CONCLUSIONS: Radiographically severe (grade III or IV) HO after CDA with the Prestige LP disc at 2 levels did not significantly affect efficacy or safety outcomes (severe AEs or secondary surgeries). However, severe HO, particularly grade IV HO, significantly limited ROM, as expected.