Twelve-Month Follow-up of Lumbar Spine Range of Motion Following Intervertebral Disc Replacement Using Radiostereometric Analysis.
Publication/Presentation Date
1-1-2008
Abstract
BACKGROUND: Many clinical studies have focused on clinical pain scores and less on kinematics following intervertebral disc replacement. Although flexion and extension of the motion segment can be measured on lateral X-rays, measuring lateral bending and axial rotation of the device is extremely difficult on plain radiography. This study was designed to measure, using radiostereometric analysis (RSA), the postoperative range of motion of the spinal segment following placement of ProDisc-L interbody device (Synthes Spine, West Chester, Pennsylvania).
METHODS: Twelve patients (15 discs) with a ProDisc-L intervertebral disc replacement were followed postoperatively at 1.5, 3, 6, and 12 months with both clinical and RSA examinations. For follow-up RSA analysis, 4 to 5 tantalum beads were inserted into the vertebrae adjacent to the surgical level during surgery. Standing biplanar films were collected during follow-up, and the ranges of motion (ROM) (sagittal and coronal bending) of the adjacent vertebrae were determined by RSA.
RESULTS: Based on the clinical surveys, this group of patients had similar outcomes compared to larger clinical populations. The flexion/extension ROM with the disc replacement averaged 2.5° at 6 weeks and increased over the follow-up period to 6.6° at 6 months. The lateral bending ROM with the disc replacement remained consistent over the 4 time points and averaged 3.0°. The motion at the level of the L4-5 vertebrae following disc replacement was greater across all time points than the motion at the L5-S1 level for both sagittal (5.9° versus 2.1°) and coronal (4.2° versus 0.6°) bending.
CONCLUSIONS: In this study, the amount of RSA-measured segmental flexion/extension ROM for those with disc replacement was similar to other studies using plain radiography. In lateral bending, the amount of motion with disc replacement was less than the typical 6°-16° reported for normal ROM.
CLINICAL RELEVANCE: This is the first published study evaluating the in vivo kinematics of artificial disc replacement using RSA.
Volume
2
Issue
1
First Page
9
Last Page
15
ISSN
1935-9810
Published In/Presented At
Ordway, N. R., Fayyazi, A. H., Abjornson, C., Calabrese, J., Park, S. A., Fredrickson, B., Yonemura, K., & Yuan, H. A. (2008). Twelve-Month Follow-up of Lumbar Spine Range of Motion Following Intervertebral Disc Replacement Using Radiostereometric Analysis. SAS journal, 2(1), 9–15. https://doi.org/10.1016/SASJ-2007-0115-RR
Disciplines
Medicine and Health Sciences
PubMedID
25802596
Department(s)
Department of Surgery
Document Type