Simple revision of the Denver peritoneovenous shunt with nonfunction from inflow occlusion.
Publication/Presentation Date
7-1-1987
Abstract
Occlusion of the Denver peritoneovenous shunt can occur when fibrinous debris block the inflow tubing or valve chamber. In this situation, revision is simplified by replacing only the valve chamber and inflow tubing, without requiring exposure of the central vein or the peritoneum. The outflow tubings are then joined over a Teflon adapter.
Volume
165
Issue
1
First Page
71
Last Page
72
ISSN
0039-6087
Published In/Presented At
Moritz, M. J., Hoch, J. R., & Rosato, F. E. (1987). Simple revision of the Denver peritoneovenous shunt with nonfunction from inflow occlusion. Surgery, gynecology & obstetrics, 165(1), 71–72.
Disciplines
Medicine and Health Sciences
PubMedID
3589931
Department(s)
Department of Surgery
Document Type