Multicenter Evaluation of a New Strip-Based Blood Glucose System for Point-of-Care Testing in Critical and Non-Critical Care Settings.

Publication/Presentation Date

7-1-2026

Abstract

BACKGROUND: Evaluation of the performance of Cobas

METHODS: Arterial, capillary, venous, or heel stick whole blood (WB) samples from patients in non-critical and critical care settings were collected and analyzed using BGMSA and the Nova StatStrip

RESULTS: Two studies are presented. In the first, 2678 samples (622 arterial, 706 capillary, 1203 venous, 147 heel stick) were collected from 1577 patients in 14 US and three European sites. All accuracy criteria were met for arterial and venous samples considering all data combined. BGMSA showed better accuracy than BGMSB for arterial, venous, and heel stick, and similar results to BGMSB for capillary WB vs venous comparator. No endogenous interference from pO

CONCLUSIONS: BGMSA was effective for determining blood glucose in venous, arterial, neonatal arterial, neonatal heel stick, and capillary WB samples.