The Performance Characteristics of Handheld, Non-Piezoelectric Point-of-Care Ultrasound (POCUS) in the Emergency Department.

Publication/Presentation Date

12-21-2023

Abstract

The use of handheld point-of-care ultrasound (HH-POCUS) platforms is rapidly increasing, but the diagnostic performance of HH-POCUS in the emergency department (ED) has not been well-studied. For a period of one year, only a HH-POCUS system that uses a non-piezoelectric array (Butterfly iQ+™) was available for clinical POCUS examinations in our ED. We performed a retrospective observational study of patients who underwent cardiac, thoracic, renal, biliary, or lower extremity venous (DVT) examinations from November 2021-November 2022 and calculated performance characteristics of HH-POCUS relative to radiology imaging. A total of 381 HH-POCUS studies were evaluated. Cardiac image quality was significantly lower than lung (

Volume

14

Issue

1

ISSN

2075-4418

Disciplines

Medicine and Health Sciences

PubMedID

38201326

Department(s)

Fellows and Residents

Document Type

Article

Share

COinS