Cumulative radiation dose from medical imaging in paediatric congenital heart disease patients with epicardial cardiac implantable electronic devices.

Publication/Presentation Date

1-1-2024

Abstract

AIMS: To determine whether paediatric congenital heart disease (CHD) patients with epicardial cardiac implantable electronic devices (CIEDs) receive high cumulative effective doses (CEDs) of ionizing radiation from medical imaging tests.

METHODS AND RESULTS: We compared 28 paediatric CHD patients with epicardial CIEDs (cases) against 40 patients with no CIED matched by age at operation, sex, surgical era, and CHD diagnosis (controls). We performed a retrospective review of radiation exposure from medical imaging exams between 2006 and 2022. Radiation dose from computed tomography (CT) and X-ray radiography was calculated using the National Cancer Institute Radiation Dosimetry Tool. We performed univariate analysis to compare the CED between the two groups. In the case subgroup, we convened experts' review to adjudicate the prevalence of CT exams that should have been performed with magnetic resonance imaging (MRI) in the absence of a CIED. Children (median age 2.5 years at implant) with CIEDs received significantly higher median CED compared with matched controls (6.90 vs. 1.72 mSv,

CONCLUSION: Paediatric CHD patients with CIED received four times higher CED than matched controls. Improved access to medical imaging tests without ionizing radiation, such as MRI, could potentially reduce the ED in CIED patients by up to five times.

Volume

2

Issue

1

First Page

060

Last Page

060

ISSN

2755-9637

Disciplines

Medicine and Health Sciences

PubMedID

39045197

Department(s)

Fellows and Residents

Document Type

Article

Share

COinS