Management of arch aneurysms with a single-branch thoracic endograft in zone 0.

Publication/Presentation Date

6-1-2021

Abstract

BACKGROUND: We present preliminary data from a patient cohort undergoing thoracic endovascular aortic repair for Ishimaru zone 0 and 1 using a novel branched arch endograft.

METHODS: This US multicenter early feasibility investigational device exemption clinical trial treated 9 patients with a mean age 72.8 ± 8.0 years (77.8% male). The endograft was designed with a single side branch designed to facilitate aortic coverage proximal to the innominate or left carotid artery while maintaining branch vessel patency. Pathology treated included fusiform (n = 2) or saccular (n = 7) aneurysm, with a maximum aortic diameter of 6.3 ± 0.7 cm. Treatment was into zone 0 in 8 patients, and zone 1 in 1 patient.

RESULTS: All patients underwent initial successful first-stage supra-aortic trunk revascularization using a variety of techniques, without the occurrence of stroke. For the second thoracic endovascular aortic repair stage, median total treatment length was 20 cm. The primary end point of device delivery and branch vessel patency was achieved in 100% of patients, without 30-day mortality or spinal cord ischemia. Cerebrovascular events were observed in 2 patients through 30 days. No type I or III endoleaks were reported and all side branches were patent at 12-month imaging follow-up.

CONCLUSIONS: Endovascular repair of Ishimaru zone 0 or 1 arch aortic aneurysms can be achieved with a novel branched arch endograft. Future studies will evaluate the mid-term outcomes with this device in other pathologies and further define the occurrence of postoperative neurologic events.

Volume

7

First Page

1

Last Page

6

ISSN

2666-2507

Disciplines

Medicine and Health Sciences

PubMedID

34318189