Effect of Phototherapy on the Reliability of Transcutaneous Bilirubin Devices in Term and Near-Term Infants: A Systematic Review and Meta-Analysis.
Publication/Presentation Date
1-1-2016
Abstract
BACKGROUND: Transcutaneous bilirubin (TcB) devices are commonly used for screening of hyperbilirubinemia in term and near-term infants not exposed to phototherapy. However, the accuracy of TcB devices in infants exposed to phototherapy is unclear.
OBJECTIVES: To conduct a systematic review of studies comparing TcB devices with total serum bilirubin (TSB) in infants receiving phototherapy or in the postphototherapy phase.
METHODS: MEDLINE, EMBASE, Cochrane Library, CINAHL and Scopus databases (from inception to May 8, 2014) were searched. Additional citations were identified from the bibliography of selected articles and from the abstracts of conference proceedings. The studies were included if they compared TcB results with TSB in term and near-term infants during phototherapy or after discontinuation of phototherapy. Two reviewers independently assessed studies for inclusion, and discrepancies were resolved with consensus. Risk of bias was assessed using the QUADAS-2 tool.
RESULTS: Fourteen studies were identified. The pooled estimates of correlation coefficients (r) during phototherapy were: covered sites 0.71 (95% CI 0.64-0.77, 11 studies), uncovered sites 0.65 (95% CI 0.55-0.74), 8 studies), forehead 0.70 (95% CI 0.64-0.75, 12 studies) and sternum 0.64 (95% CI 0.43-0.77, 5 studies). Two studies also provided results as Bland-Altman difference plots (mean TcB-TSB differences -29.2 and 30 µmol/l, respectively). The correlation coefficient improved marginally in the postphototherapy phase (r = 0.72, 95% CI 0.64-0.78, 4 studies).
CONCLUSION: We found a moderate correlation between TcB and TSB during phototherapy with a marginal improvement in the postphototherapy phase. Further research is needed before the use of TcB devices can be recommended for these settings.
Volume
109
Issue
3
First Page
203
Last Page
212
ISSN
1661-7819
Published In/Presented At
Nagar, G., Vandermeer, B., Campbell, S., & Kumar, M. (2016). Effect of Phototherapy on the Reliability of Transcutaneous Bilirubin Devices in Term and Near-Term Infants: A Systematic Review and Meta-Analysis. Neonatology, 109(3), 203–212. https://doi.org/10.1159/000442195
Disciplines
Medicine and Health Sciences | Pediatrics
PubMedID
26789390
Department(s)
Department of Pediatrics
Document Type
Article